Influence of Thermistor Probe Depth from the Anterior Nares on Measurement of Nasopharyngeal Temperature
Hiroaki Sato1, Michiaki Yamakage2, Katsumi Okuyama1, Yusuke Imai1, Hironobu Iwashita1, Taishi Masamune1, Tadahiko Ishimaya1 and Takashi Matsukawa1
1Department of Anesthesiology, University of Yamanashi, Faculty of Medicine, Chuo, Yamanashi, Japan. 2Department of Anesthesiology, Sapporo Medical University School of Medicine, Sapporo, Hokkaido, Japan.
Abstract
Nasopharyngeal temperature is allegedly accurate and is generally used during cardiopulmonary bypass for open-heart surgery. However, adequate depth from the anterior nares to measure nasopharyngeal temperature has not been evaluated. To test whether nasopharyngeal temperature is sufficiently accurate and precise for clinical use and to clarify the suitable depth of insertion, we compared nasopharyngeal temperature measurements to simultaneous tympanic temperature measurements during open-heart surgery with cardiopulmonary bypass. Subjects comprised 4 women and 6 men undergoing cardiac surgery with a target core temperature of 32 °C. Nasopharyngeal temperature was measured at 4 sites by placing thermocouples in the nasal cavity in 1-cm increments starting at 2 cm from the anterior nares. The reference temperature (tympanic temperature) was measured at the right tympanic membrane using a thermocouple. Both temperatures were measured every 5 min and compared using correlation coefficients of linear regression (r2) and bias (mean difference between the two methods). Compared to tympanic temperature, nasopharyngeal temperature showed regression slopes of 0.91–1.03 and correlation coefficients of 0.55–0.65 at all 4 depths. Accuracy (offset or bias) was 0.9–2.1 °C compared to tympanic temperature. Precision (standard deviation) of measurements was 0.8–1.2 °C. Nasopharyngeal temperature collected at 2 cm varied the most from tympanic temperature. In conclusion, the reliability of T Naso is low for monitoring core body temperature during open-heart surgery with CPB, particularly if the probe is placed at a shallow depth from the anterior nares. Nasopharyngeal temperature obtained closest to the anterior nares (2 cm) were the least accurate.
Readers of this also read:
- A Measuring Medical Pocket Calculator
- A Simple Derivation of the Distribution of Pairwise Local Protein Sequence Alignment Scores
- Evaluation of Two Outlier-Detection-Based Methods for Detecting Tissue-Selective Genes from Microarray Data
- DICOM Structured Reporting and Cancer Clinical Trials Results
- Mammoth and Elephant Phylogenetic Relationships: Mammut Americanum, the Missing Outgroup